H₂O₂ Low Temperature Plasma Sterilizer Use
Product Description
Product Detail
Characteristics
◆We exclusively adopt the intelligent high-precision H₂O2 injection system.Through theintelligent closed-loop feedback system, the injection volume and loading volume are precisely controlled.That saves a lot of consumables,and minimizes hospital costs.
◆ The unique sterilant purification technology enables hydrogen peroxide to diffuse more evenly and fully,especially when sterilizing narrow and long tube,and this core technology has been patented.
◆Excellent sterilization performance.For the stainless steel tube,with an inner diameter of 1mm,it can effectively sterilize over a distance of 500mm.Additionally,for the polytetrafluoroethylene hose with 1mm inner diameter,delivers outstanding sterilization results across 2,000mm.
◆Electric push rodtype lifting door,equipped with infrared control safety system,is able to perceive and feedback potential obstacle in advance without touching,which is safe and reliable.
◆The single multi-frequency alternating current discharge mode ensures 100%successful plasma discharge.The ultraviolet rays generated by the plasma are completely shielded to ensure the operators are not affected by radiation.
◆ The core components are all sourced from international renowned brands,ensuring a low failure rate for the entire unit.
◆The friendly HMI displays all parameters of the entire process,making the operation both convenient and flexible.
◆ It supports intelligent remote interconnection,and enables remote fault diagnosis,remote maintenance,remote training,etc.
◆Power supply 380V/50hz,it is with casters at the bottom,convenient to more.
◆ The sterilizing chamber is made of 5052 anti-rust alloy,which has good heat conduction performance,ensuring the temperature in the sterilization cabin remains balanced.
◆ The lower sterilizing temperature,50℃(±5℃),is harmlessto the instrument and beneficial for prolonging its usage life and reducing maintenance costs.
◆This unit has obtained multiple patent certifications and is officially registered with the CFDA.
Technical Data
|
MODEL/Technical Data |
BJ-PS80 |
BJ-PS120 |
BJ-PS200 |
|
Chamber shape |
|||
|
Rectangular shape |
Rectangular shape |
Rectangular shape |
|
|
Overall size(W×H×D) |
746×1700×940mm |
845×1700×990mm |
932×1775×1098mm |
|
Chamber size(W×H×D) |
400×360×680mm |
450×410×780mm |
550×445×900mm |
|
Chamber capacity |
98L |
144L |
220L |
|
Valid capacity |
80L |
120L |
200L |
|
Built-in sterilization programs |
Fast cycle,standard cycle,enhanced cycle |
||
|
Sterilization result |
Stainles stel tubes with inner diameter of 1mm and length of 500mm;polytetrafluoroethylene tubes with inner diameter of lmm and length of 2000mm |
||
|
Standard sterilizing cycle time |
≤45min |
<50min |
≤55min |
|
Control system |
Fully automaticmicrocomputer control,real time monitoring,automatic preheat,sarup detection,display the malfunction info on the screen along with sound alarm. |
||
|
Sterilizing temperature |
50±5℃ |
||
|
Sterilant |
56%to 60%medical-grade hydrogen peroxide |
||
|
Sterilant usage |
Single cycle,each time 1.7ml |
Single cycle,each time 2.5ml |
Single cycle,each time 3.8ml |
|
Injection accuracy |
1% |
||
|
Operation interface |
English operation interface,for setting or print the sterilizing parameter of the full process. |
||
|
Recording method |
The built-in printer can print the parameters of each sterilization result in real time,including sterilization time,temperature,etc. |
||
|
Safety device |
Infrared anti-pinch hand protection |
||
|
Imported components |
Breaker,contactor,solenoid valve,relay,photoelectric switch,microswitch |
||
|
Door open/close |
The electric push rod enables automatic lifing and descending. |
||
|
Minimum installation space |
1500×1200mm |
||
|
Minimum repairing space |
1000mm |
||
|
Minimum distance from the rear shell to the wall |
150mm |
||
|
Installation environment |
10-40℃ |
||
|
Power supply |
AC380V/50Hz |
||
|
Consumption |
2.8KW |
3KW |
3.5KW |
|
Weight |
275kg |
340kg |
390kg |